At a glance:
- Transmittance dust measuring devices DUSTHUNTER T100
- For medium to high dust concentrations
- On-board contamination check
- Automated thorough check of zero and reference point
- For small to large measuring distances
Please note: The exact device specifications and performance data of the product may deviate from the information provided here, and depend on the application in which the product is being used and the relevant customer specifications.
Technical data
DUSTHUNTER T100 system
- Measured values: Transmittance, opacity, relative opacity, extinction, dust concentration
- Performance-tested measurands: Dust concentration
- Measurement principles: Transmittance measurement
- Spectral range : 450 nm … 700 nm
- Measuring ranges:
Transmittance: 100 … 80 % / 100 … 0 %
Opacity: 0 … 20 % / 0 … 100 %
Relative opacity: 0 … 20 % / 0 … 100 %
Extinction: 0 … 0.1 / 0 … 2
Dust concentration: 0 … 200 mg/m³ / 0 … 10,000 mg/m³
The measurement depends on measuring distance and dust properties
- Certified measuring ranges: Dust concentration 0 … 0.1 Ext / 0 … 0.05 Ext / 0 … 0.2 Ext / 0 … 0.5 Ext / 0 … 1 Ext (transmittance)
- Response time (t90): 1 s … 600 s, Freely adjustable
- Accuracy: ± 2 %
- Process temperature: –40 °C … +600 °C
- Process pressure:
With MCU-P control unit –50 hPa … 2 hPa
With external purge air unit –50 hPa … 30 hPa
- Process gas humidity: Non-condensing
- Duct diameter: 0.5 m … 2.5 m/ 2 m … 5 m/ 4 m … 12 m
- Conformities: Approved for plants requiring approval
2001/80/EC (13. BImSchV)
2000/76/EC (17. BImSchV)
27. BImSchV
TA-Luft (Prevention of Air Pollution)
EN 15267
EN 14181
MCERTS
2010/75/EU
- Electrical safety: CE
- Test functions: Automatic self-test (linearity, contamination, drift, aging)/ Contamination limit values: warning at 20%, fault at 30%/ Manual linearity test with reference filter
- Options: External purge air unit