DUSTHUNTER SB30 system
- Measured values: Scattered light intensity, dust concentration (after gravimetric comparison measurement)
- Measurement principles: Scattered light forward
- Spectral range: 640 nm … 660 nm, Laser, protection class 2, power < 1 mW
- Measuring ranges:
Scattered light intensity: 0 … 7.5 SI / 0 … 3,000 SI
Measuring ranges freely selectable; nine measuring ranges pre-configured (0 … 7.5/15/45/75/150/225/375/1,000/3,000 SI)
- Response time (t90): 0.1 s … 600 s, Freely adjustable via SOPAS ET software
- Accuracy: ± 2 %, Of measuring range full scale
- Process temperature: –40 °C … +220 °C
- Process pressure:
With intergrated purge air unit: –50 hPa … 10 hPa
With external purge air unit: –50 hPa … 30 hPa
With instrument air (provided by the customer): –50 hPa … 100 hPa
Process gas humidity
Non-condensing
- Duct diameter:
Tri-Clamp version, 1″ thread version: ≥ 150 mm
Flange version: ≥ 250 mm
- Conformities: TUEV type-examination
- Electrical safety: CE
- Test functions:
Automatic self-test (linearity, drift, aging)
Manual linearity test with reference filter
- Options:
MCU control unit
Integrated purge air unit
External purge air unit
DHSP30 sender/receiver unit
- Ambient temperature: –40 °C … +60 °C
- Electrical safety: CE
- Enclosure rating: IP65
- Analog outputs: 1 output, 0/4 … 20 mA, + 750 Ω, Electrically isolated
- Digital outputs: 3 relay contacts, 48 V, 1 A, For status signals
- Digital inputs: 4 inputs, For external maintenance switch, automated operational check, linearity measurement, calibration curve switching, or filter monitoring
- Communication interface:
Serial
Modbus
- Type of fieldbus integration: RTU RS-485
- Function: For connection of a MCU control unit or as service interface
- Number: 1
- Dimensions (W x H x D): Dimensions may vary. For details, see the dimensional drawings.
- Weight:
Nominal length 180 mm: ≤ 2.7 kg
Nominal length 280 mm: ≤ 2.7 kg
Nominal length 435 mm: ≤ 5 kg
Nominal length 735 mm: ≤ 6.3 kg
All weights without integrated purge air unit
- Power supply:
- Voltage: 24 V DC
- Voltage: Supply via MCU control unit or an external power supply
- Power consumption
Without integrated purge air unit: ≤ 4 W
With intergrated purge air unit: ≤ 30 W
MCU-N control unit (option)
- Description: Unit to control the system components and to evaluate and output the data provided by them
- Ambient temperature: –40 °C … +60 °C
- Electrical safety: CE
- Enclosure rating: IP65
- Analog outputs: 1 output, 0/2/4 … 20 mA, + 750 Ω, Electrically isolated; two additional outputs if using I/O modules (option)
- Analog inputs: 2 inputs, 0 … 20 mA, Not electrically isolated; two additional inputs if using I/O modules (option)
- Digital outputs: 5 relay contacts, + 48 V, 1 A, Potential-free; for status signals
- Digital inputs: 4 potential-free contacts
- Communication interface:
Modbus
PROFIBUS DP
Ethernet
- Type of fieldbus integration:
TCP (via optional interface module, only one module possible per MCU)
RTU RS-485 (via optional interface module, only one module possible per MCU)
Via optional interface module (only one module possible per MCU)
Function: Connection to SOPAS ET software
- Number: 1
- Indication: LC display (option), Status LEDs: “Power,” “Failure,” and “Maintenance request”
- Operation: Via LC-display (option) or software SOPAS ET
- Dimensions (W x H x D): 210 mm x 340 mm x 120 mm (for details see dimensional drawings)
- Weight: ≤ 3.7 kg
- Power supply:
Voltage: 90 … 250 V
Voltage: Version with 24 V DC available as an option
Frequency: 47 … 63 Hz
Power consumption: ≤ 15 W
- Options: Interface module(s); I/O module(s)
SLV4-2 purge air unit, 2BH1300, 3-ph
- Description: Unit to provide dust-free air for flushing of optical surfaces
- Gas flow rate: 38 m³/h … 63 m³/h, At 30 hPa counter pressure, depending on low pressure inside the filter
- Ambient temperature: –40 °C … +55 °C
- Electrical safety: CE
- Enclosure rating: IP54
- Dimensions (W x H x D): 550 mm x 550 mm x 257 mm (for details see dimensional drawings)
- Weight: 18 kg
- Power supply: Three-phase current
Δ: 200 … 240 V, 50 Hz, 2,6 A, 400 W
Y: 345 … 415 V, 50 Hz, 1,5 A, 400 W
Δ: 200 … 275 V, 60 Hz, 2,6 A, 500 W
Y: 380 … 480 V, 60 Hz, 1,5 A, 500 W
Δ: 270 … 330 V, 50 Hz, 2,0 A, 400 W
Y: 465 … 570 V, 50 Hz, 1,16 A, 400 W
Δ: 290 … 360 V, 60 Hz, 2,1 A, 500 W
Y: 500 … 600 V, 60 Hz, 1,26 A, 500 W
Δ: 230 V, 50 Hz, 2,7 A, 370 W
Δ: 115 V, 60 Hz, 3,0 A, 450 W
Δ: 220 … 270 V, 50 Hz, 2,5 A, 400 W
Y: 380 … 465 V, 50 Hz, 1,45 A, 400 W
Δ: 240 … 290 V, 60 Hz, 2,6 A, 500 W
Y: 415 … 500 V, 60 Hz, 1,55 A, 500 W
- Auxiliary gas connections: Purge air 40 mm
- Test functions
Pressure switch (switching point –35 hPa)
Integrated components: 2-step air filter, type Europiclon, dust capacity 200 g