At a glance:
- Extractive gas analyzers S700
- Measured values: Ar, CBrF3, CClF3, CCl2F2, CCl3F, CHClF2, CHCl3, CH2Cl2, CH2O, CH3Cl, CH3OH, (CH3)2O, CH4, CO, CO2, CS2, C2Cl2F4, C2Cl3F3, C2Cl4, C2HCl3, C2H2, C2H2Cl2, C2H2F4, C2H3Cl3, C2H4, C2H4Cl2, C2H5OH, C2H6, C3H4, C3H6, C3H7OH, C3H8, C3H8O2, C4H6, C4H8, C4H10, C4H8O, C4H9OH, C5H12, C6H10O, C6H12, C6H14, C7H16, C7H8, C8H10, He, H2, H2O, NH3, NO, N2O, O2, SF6, SO2
- 3 different housing variants for different fields of application
- Up to 3 analyzer modules in a single housing
- Easy application-specific adaptation due to modular design
- Also suitable for zone 1 and zone 2 explosive environments (ATEX)
- Automatic adjustment with test gas or calibration cell
- Integrated self-monitoring and fault diagnosis
- Emission measurement according to 13th (2001/80/EC), 17th (2000/76/EC), and 27th German Federal Emission Protection Directive (BImSchV).
- Cooling gas monitoring of turbo generators
- Measurement of CO for coal mill monitoring
- Purity measurement of H2 in pressure swing adsorption plants
- Monitoring of CO2 in natural gas in natural gas conditioning plants
Please note: The exact device specifications and performance data of the product may deviate from the information provided here, and depend on the application in which the product is being used and the relevant customer specifications
Technical data:
S700 system
- Measured values: Ar, CBrF3, CClF3, CCl2F2, CCl3F, CHClF2, CHCl3, CH2Cl2, CH2O, CH3Cl, CH3OH, (CH3)2O, CH4, CO, CO2, CS2, C2Cl2F4, C2Cl3F3, C2Cl4, C2HCl3, C2H2, C2H2Cl2, C2H2F4, C2H3Cl3, C2H4, C2H4Cl2, C2H5OH, C2H6, C3H4, C3H6, C3H7OH, C3H8, C3H8O2, C4H6, C4H8, C4H10, C4H8O, C4H9OH, C5H12, C6H10O, C6H12, C6H14, C7H16, C7H8, C8H10, He, H2, H2O, NH3, NO, N2O, O2, SF6, SO2
- Performance-tested measurands: CO, NO, O2, SO2
- Gas flow rate:
No integrated sample gas pump: 5 l/h … 100 l/h
With integrated sample gas pump: 30 l/h … 60 l/h
- Sample gas temperature: 0 °C … +45 °C, Temperature at analyzer inlet
- Process pressure:
Tubed gas lines: –200 hPa … 1,000 hPa
Hosed gas lines: –200 hPa … 300 hPa
- Process gas humidity: Non-condensing
- Dust load: Free of dust and aerosols
- Ambient temperature: +5 °C … +45 °C
- Storage temperature: –20 °C … +70 °C
- Ambient pressure: 700 hPa … 1,200 hPa
- Geographical altitude: ≤ 2,000 m (above mean sea level)
- Ambient humidity: ≤ 95 %, Relative humidity; non-condensing
- Electrical safety: CE, cCSAus
- Analog outputs: 4 outputs, 0/4 … 20 mA, 500 Ω, 0 … 10 V, Electrically isolated
- Analog inputs: 2 inputs, 0/2/4 … 20 mA (Option: 0 … 10 V DC)
- Digital outputs :
8 relay contacts: Three relay outputs preset for failure, service and maintenance
8 Open collector outputs: Freely adjustable
- Digital inputs: 8 optical coupler inputs: Electrically isolated; freely programmable
- Communication interface: Modbus
- Type of fieldbus integration: RTU RS-232
- Number: 1
- Indication: LC display
- Operation: Menu-driven operation via LC-display and membrane keyboard
- Menu languages: German, English, French, Italian, Dutch, Polish, Swedish, Spanish
- Power supply:
Voltage: 100 V AC / 115 V AC / 230 V AC
Frequency: 48 … 62 Hz
Power consumption: ≤ 150 W (Depending on system configuration)
- Corrective functions: Automatic testing and adjustment with test gases/ Manual adjustment with test gases
- Options:
Integrated sample gas pump (only with hosed gas lines)
Tubed gas lines
Up tu three separate gas lines
Flow sensor
Humidity sensor
Barometric pressure correction
Sample gas pressure correction
Sample point switching (max. 8 sample points)
Dimensions:
Connection diagram: