At a glance:
- CEMS solutions MCS100FT
- The FTIR measuring principle allows the simultaneous determination of more than 12 measuring components – tailor-made for the particular requirements of the customer.
- Lowest approved HF measuring range: 0 … 3 mg/m³
- Automatic spectrum adjustment via AutoVAL for reliable measuring values
- Operation via touchscreen
- Sample gas transport by an ejector without moving parts
- Approved according to EN 15267
- Remote control and diagnosis via software SOPAS ET
- Automatic adjustment of analyzer
- Automatic backflushing and filter cleaning of sampling unit
- Emission monitoring in waste incineration plants
- Effective HF limit value monitoring, e. g., in aluminum production
- Monitoring of gaseous emissions in cement plants or power stations
Please note: The exact device specifications and performance data of the product may deviate from the information provided here, and depend on the application in which the product is being used and the relevant customer specifications
Technical data:
MCS100FT system
- Measured values: CH4, CO, CO2, Corg, HCl, HF, H2O, NH3, NO, NO2, N2O, O2, SO2, NOx, C3H8, C2H6
- Performance-tested measurands: CH4, CO, CO2, Corg, HCl, HF, H2O, NH3, NO, NO2, N2O, O2, SO2
- Measurement principles: FTIR spectroscopy, flame ionisation detection, Zirconium dioxide sensor
- Gas flow rate: ≤ 300 l/h
- Measuring ranges:
CH4: 0 … 70 ppm / 0 … 210 ppm
CO: 0 … 60 ppm / 0 … 1,200 ppm
CO2: 0 … 25 Vol.-%
Corg: 0 … 7.5 ppm / 0 … 75 ppm
HCl: 0 … 10 ppm / 0 … 100 ppm
HF: 0 … 3 ppm / 0 … 10 ppm
H2O: 0 … 40 Vol.-%
NH3: 0 … 13 ppm / 0 … 65 ppm
NO: 0 … 150 ppm / 0 … 1,500 ppm
NO2: 0 … 25 ppm / 0 … 250 ppm
N2O: 0 … 25 ppm / 0 … 250 ppm
O2: 0 … 21 Vol.-%
SO2: 0 … 25 ppm / 0 … 525 ppm
NOx: 0 … 100 ppm / 0 … 1,000 ppm
C3H8: 0 … 25 ppm
C2H6: 0 … 40 ppm
Other measuring ranges and components on request
- Certified measuring ranges:
CH4: 0 … 50 mg/m³ / 0 … 150 mg/m³
CO: 0 … 75 mg/m³ / 0 … 300 mg/m³ / 0 … 1,500 mg/m³
CO2: 0 … 25 Vol.-%
Corg: 0 … 15 mg/m³ / 0 … 50 mg/m³ / 0 … 150 mg/m³ / 0 … 500 mg/m³
HCl: 0 … 15 mg/m³ / 0 … 90 mg/m³ / 0 … 150 mg/m³
HF: 0 … 3 mg/m³ / 0 … 10 mg/m³
H2O: 0 … 40 Vol.-%
NH3: 0 … 10 mg/m³ / 0 … 50 mg/m³
NO: 0 … 200 mg/m³ / 0 … 400 mg/m³ / 0 … 2,000 mg/m³
NO2: 0 … 100 mg/m³ / 0 … 500 mg/m³
N2O: 0 … 50 mg/m³ / 0 … 500 mg/m³
O2: 0 … 21 Vol.-%
SO2: 0 … 75 mg/m³ / 0 … 300 mg/m³ / 0 … 1,500 mg/m³
- Response time (t90): ≤ 200 s
FID: ≤ 45 s
- Sensitivity drift : < 3 % of the measuring range full scale value per maintenance interval
FID: < 2 % of measuring range full scale per week
- Zero point drift: < 3 % of the measuring range full scale value per maintenance interval
FID: < 2 % of measuring range full scale per week