MCS100E HW system
- Measured values: CH4, CO, CO2, H2O, HCl, N2O, NH3, NO, NO2, O2, SO2
- Performance-tested measurands: CO, CO2, H2O, HCl, NH3, NO, O2, SO2
- Measurement principles: Interference filter correlation, Gas filter correlation, Zirconium dioxide sensor
- Gas flow rate: 200 l/h … 1,000 l/h (Depending on application)
- Measuring ranges:
CH4: 0 … 70 ppm / 0 … 1,400 ppm
CO: 0 … 60 ppm / 0 … 2,000 ppm
CO2: 0 … 25 Vol.-% / 0 … 100 Vol.-%
H2O: 0 … 1 Vol.-% / 0 … 40 Vol.-%
HCl: 0 … 10 ppm / 0 … 1,900 ppm
NH3: 0 … 30 ppm / 0 … 660 ppm
NO: 0 … 150 ppm / 0 … 1,900 ppm
NO2: 0 … 50 ppm / 0 … 500 ppm
N2O: 0 … 50 ppm / 0 … 1,000 ppm
O2: 0 … 1 Vol.-% / 0 … 21 Vol.-%
SO2: 0 … 25 ppm / 0 … 5,000 ppm
Other measuring ranges on request
- Certified measuring ranges:
CO: 0 … 75 mg/m³
CO2: 0 … 20 Vol.-%
H2O: 0 … 40 Vol.-%
HCl: 0 … 15 mg/m³
NH3: 0 … 20 mg/m³
NO: 0 … 200 mg/m³
O2: 0 … 21 Vol.-%
SO2: 0 … 75 mg/m³
- Response time (t90): ≤ 200 s, Depending on application
- Process temperature: 0 °C … +1,300 °C
- Sample gas temperature: Inlet analyzer system 0 °C … +220 °C
- Process pressure: 900 hPa … 1,100 hPa/ Atmospheric
- Ambient temperature: +5 °C … +35 °C
- With cooling device: +5 °C … +50 °C
- Ambient pressure: 900 hPa … 1,100 hPa
- Conformities: Approved for plants requiring approval
2001/80/EC (13. BImSchV)
2000/76/EC (17. BImSchV)
27. BImSchV
TA-Luft (Prevention of Air Pollution)
EN 15267 (MCERTS)
EN 14181:2004
- Electrical safety: CE
- Enclosure rating: IP43
- Analog outputs: 4 outputs, 0/4 … 20 mA, + 500 Ω, Max. 64 outputs with opto box via optical fibre possible
- Analog inputs 4 inputs: 0/4 … 20 mA, Max. 64 inputs with opto box via optical fibre possible
- Digital outputs: 12 outputs, + 40 V, 1 A, Max. 64 outputs with opto box via optical fibre possible
- Digital inputs: 4 inputs, + 50 V AC, 4 A / 24 V DC, 4 A, Max. 64 inputs with opto box via optical fibre possible
- Communication interface: Modbus/ PROFIBUS DP
- Type of fieldbus integration: RTU RS-232/ RTU RS-485/ TCP
- Number: 1
- Operation:
Via integrated control unit
Two operating levels, one password-protected
Sequence programs can be programmed as required
- Dimensions (W x H x D): 806 mm x 2,165 mm x 605 mm (for details see dimensional drawings)
- Test functions: Automatic check cycle for zero and span point/ Internal calibration filter for QAL3 drift check without test gas (option)
- System components:
Analyzer cabinet
Sampling probe
Sample gas line, heated
- Options:
Zirconium dioxide sensor (oxygen measurement)
Total hydrocarbon analyzer