ZIRKOR200 analyzer unit standard version
- Measured values: O2
- Performance-tested measurands: O2
- Measurement principles: Zirconium dioxide sensor
- Measuring ranges: 0 … 25 Vol.-%
- Certified measuring ranges:
O2: 0 … 25 Vol.-%
- Response time (t90): 5 s, Depending on sample gas flow
- Accuracy: ± 0.5 %, Of the measured value or 0.02% by O2 volume, with the higher value being valid
- Process temperature:
- Measuring probe: ≤ +600 °C
- Measuring probe with protective tube: ≤ +1,600 °C
- Process pressure: –100 hPa … 100 hPa, Maximum pressure fluctuation: ± 50 hPa
- Process gas velocity: ≤ 50 m/s
- Ambient temperature: –40 °C … +80 °C
- Conformities:
EN 15267
2000/76/EC (17. BImSchV)
2001/80/EC (13. BImSchV)
27. BImSchV
30. BImSchV
TA-Luft (Prevention of Air Pollution)
- SIL:
Safety integrity level: SIL 2 (IEC 61508)
Systematic capability: SC 2 (IEC 61508)
- Device type: Type B
- Operating mode: Low demand
- System architecture: 1oo1 voting
- Comment: Option
- Electrical safety: CE
- Enclosure rating: IP65
- Dimensions (W x H x D): Dimensions may vary. For details, see the dimensional drawings.
- Immersion depth:
Measuring probe: ≤ 3,682 mm
Measuring probe with protective tube: ≤ 1,000 mm
- Material in contact with media: Stainless steel 1.4571
- Material: Stainless steel 1.4571
- Options:
Protective tube
LongLife cell
LongLife² cell
ZIRKOR200 analyzer unit Ex versions
- Measured values: O2
- Performance-tested measurands: O2
- Measurement principles: Zirconium dioxide sensor
- Measuring ranges: 0 … 25 Vol.-%
- Certified measuring ranges:
O2: 0 … 25 Vol.-%
- Response time (t90): 5 s, Depending on sample gas flow
- Accuracy: ± 0.5 %, Of the measured value or 0.02% by O2 volume, with the higher value being valid
- Process temperature:
ZIRKOR200 Ex-G measuring probe: ≤ +500 °C
ZIRKOR200 Ex-G measuring probe with cooling protection tube: ≤ +1,400 °C
ZIRKOR200 Ex-D measuring probe: ≤ +600 °C
- Process pressure: –100 hPa … 100 hPa, Maximum pressure fluctuation: ± 50 hPa
- Process gas velocity: ≤ 50 m/s
- Ambient temperature:
ZIRKOR200 Ex-G: –20 °C … +55 °C
ZIRKOR200 Ex-D: –20 °C … +70 °C
- Conformities:
EN 15267
2000/76/EC (17. BImSchV)
2001/80/EC (13. BImSchV)
27. BImSchV
30. BImSchV
TA-Luft (Prevention of Air Pollution)
- SIL:
Safety integrity level: SIL 2 (IEC 61508)
Systematic capability: SC 2 (IEC 61508)
- Device type: Type B
- Operating mode: Low demand
- System architecture: 1oo1 voting
- Comment: Option
- Ex-approvals:
IECEx: ZIRKOR200 Ex-G: Ex db IIC T3 Gb/ ZIRKOR200 Ex-D: Ex tb IIIC T133°C/T141°C Db
ATEX: ZIRKOR200 Ex-G: II 2G Ex db IIC T3 Gb/ ZIRKOR200 Ex-D: II 2D Ex tb IIIC T133°C/T141°C Db
- Electrical safety: CE
- Enclosure rating: ZIRKOR200 Ex-G, IP66/ ZIRKOR200 Ex-D
- Dimensions (W x H x D): Dimensions may vary. For details, see the dimensional drawings.
- Immersion depth:
ZIRKOR200 Ex-G measuring probe: ≤ 924 mm
ZIRKOR200 Ex-D measuring probe: ≤ 960 mm
- Material in contact with media: Stainless steel 1.4571
- Material: Stainless steel 1.4571
- Options:
Cooling protection tube (not with ZIRKOR200 Ex-D)
LongLife cell (already included with ZIRKOR200 Ex-D)
LongLife² cell
ZIRKOR200 control unit standard version
- Ambient temperature:
- Instrument air type: –20 °C … +55 °C
- Pump type: +20 °C … +50 °C
- GRP housing with heating: –40 °C … +55 °C
- Storage temperature: –40 °C … +80 °C
- Conformities: EN 15267
- SIL:
Safety integrity level: SIL 2 (IEC 61508)
Systematic capability: SC 2 (IEC 61508)
- Device type: Type B
- Operating mode: Low demand
- System architecture: 1oo1 voting
- Comment: Option
- Electrical safety: CE
- Enclosure rating:
Steel sheet enclosure: IP66
Stainless-steel housing: IP66
GRP housing: IP66
19″-rack enclosure: IP20
- Analog outputs: 1 output, 4 … 20 mA, + 500 Ω, Electrically isolated
- Analog inputs: 1 input, 4 … 20 mA, 120 Ω, Not electrically isolated
- Digital outputs : 5 relay contacts
- Digital inputs: 2 relay contacts, + 24 V AC/DC, 1 A
- Communication interface:
Modbus
HART
Foundation Fieldbus
- Type of fieldbus integration: RTU RS-232/ RTU RS-485
- Number: 1
- Indication:
Status LEDs: “Alarm”, “Maintenance” and “Failure”
- Operation:
Via LC display and membrane keyboard
Via Android device with ZIRKOR Remote app
- Model:
Steel sheet enclosure
Stainless-steel housing
GRP housing
19″-rack
- Dimensions (W x H x D):
Steel sheet enclosure: 300 mm x 440 mm x 240 mm (for details see dimensional drawings)
Stainless-steel housing: 300 mm x 400 mm x 240 mm (for details see dimensional drawings)
GRP housing: 510 mm x 600 mm x 310 mm (for details see dimensional drawings)
19″-rack: 483 mm x 177 mm x 400 mm (for details see dimensional drawings)
- Weight:
Steel sheet enclosure: 17 kg … 19 kg
Stainless-steel housing: 17 kg … 19 kg
GRP housing: 19 kg … 27 kg
19″-rack: 10 kg … 11 kg
- Power supply:
Voltage: 104 … 126 V / 207 … 253 V
Power consumption: ≤ 400 W
Recommended pre-fuse: 10 A
- Options:
Semi-automatic and automatic adjustment (1-point or 2-point adjustment)
LongLife cell
LongLife² cell
Ejector and pump type
ZIRKOR200 control unit Ex-G version
- Ambient temperature: –20 °C … +55 °C
- Storage temperature: –40 °C … +80 °C
- Conformities: EN 15267
- SIL:
Safety integrity level: SIL 2 (IEC 61508)
Systematic capability: SC 2 (IEC 61508)
- Device type: Type B
- Operating mode: Low demand
- System architecture: 1oo1 voting
- Comment: Option
- Ex-approvals:
IECEx Ex db IIC T6 Gb
ATEX II 2G Ex db IIC T6 Gb
- Electrical safety: CE
- Enclosure rating: IP66
- Analog outputs: 1 output, 4 … 20 mA, + 500 Ω, Electrically isolated
- Analog inputs: 1 input, 4 … 20 mA, 120 Ω, Not electrically isolated
- Digital outputs : 5 relay contacts
- Digital inputs: 2 relay contacts, + 24 V AC/DC, 1 A
- Communication interface:
Modbus
HART
Foundation Fieldbus
- Type of fieldbus integration: RTU RS-232/ RTU RS-485
- Number: 1
- Indication:
Status LEDs: “Alarm”, “Maintenance” and “Failure”
- Operation:
Via LC display and membrane keyboard
Via Android device with ZIRKOR Remote app
Model: Pressure-resistant encapsulated housing
- Dimensions (W x H x D): 700 mm x 359.5 mm x 220 mm (for details see dimensional drawings)
- Weight: 32 kg
- Power supply:
Voltage: 104 … 126 V / 207 … 253 V
Power consumption: ≤ 400 W
Recommended pre-fuse: 10 A
ZIRKOR200 Ex-G pneumatic unit
- Ambient temperature: –20 °C … +55 °C
- Electrical safety: CE
- Enclosure rating: IP65
- Dimensions (W x H x D): 300 mm x 440 mm x 218 mm (for details see dimensional drawings)
- Auxiliarie:
Test gas: 150 l/h … 180 l/h
Instrument air: 30 l/h … 40 l/h
2 … 10 bar; particle size max. 1 µm; particle content max. 1 mg/m³; oil content max. 0.1 mg/m³; pressure condensation point max. –40 °C