At a glance:
- In-situ gas analyzers GM700
- The GM700 is able to measure several gas components, such as NH3, HF, HCl or H2O
- High selectivity due to high spectral resolution
- Short response times
- No calibration required
- No moving parts: minimal wear and tear
- No gas sampling or conditioning required
- Overpressure encapsulated design for ATEX Zone 2
- Ammonia slip measurement in DeNOx plants
- Emission monitoring of hydrogen fluoride in aluminum smelters
- Ammonia measurement in exhaust gas of combustion engines
- Ammonia measurement in composting or waste incineration plants
- HCI measurement in combustion plants
- HF measurement in the ceramics industry
Please note: The exact device specifications and performance data of the product may deviate from the information provided here, and depend on the application in which the product is being used and the relevant customer specifications
Technical data:
System GM700, standard version
- Description: In-situ gas analyzer for emission monitoring and process measurement
- Measured values: NH3, HF, HCl, H2O
- Performance-tested measurands: HF
- Measurement principles: Diode laser spectroscopy (TDLS)
- Measuring ranges:
NH3 (up to 20% H2O by volume): 0 … 25 ppm / 0 … 4,000 ppm
NH3 (up to 40% H2O by volume): 0 … 25 ppm / 0 … 100 ppm
HCl: 0 … 10 ppm / 0 … 3,000 ppm
HF: 0 … 5 ppm / 0 … 2,000 ppm
Measuring ranges refer to 1 m measuring path
Measuring ranges depend on application and device version
- Dual measuring ranges:
HCl / H2O: 0 … 10 ppm / 0 … 3,000 ppm (HCl)
0 … 50 Vol.-% / 0 … 100 Vol.-% (H2O)
NH3 / H2O: 0 … 25 ppm / 0 … 4,000 ppm (NH3)
0 … 20 Vol.-% / 0 … 20 Vol.-% (H2O)
NH3 / H2O: 0 … 25 ppm / 0 … 100 ppm (NH3)
0 … 40 Vol.-% / 0 … 40 Vol.-% (H2O)
Measuring ranges refer to 1 m measuring path
Measuring ranges depend on application and device version
- Certified measuring ranges:
HF: 0 … 5 mg/m³ / 0 … 25 mg/m³
Suitability-tested in the cross-duct type
- Response time (t90):
- Standard: 1 s … 360 s
- Adjustable, preset to 4 s
- HF measurement: 1 s … 180 s, Adjustable, preset to 4 s, suitability-tested according to EN 15267
- Accuracy:
Zero point: ≤ ± 2 %, Relative to measuring range end value
Sensitivity: ≤ ± 2 %, Within the maintenance interval (6 months), relative to measuring range full scale
- Ambient temperature: –20 °C … +50 °C, Depends on parameterization; temperature change max. ±10 °C/h
- Storage temperature: –40 °C … +55 °C
- Ambient humidity: ≤ 95 %, Relative humidity; non-condensing
- Conformities:
Only for HF
Approved for plants requiring approval
2001/80/EC (13. BImSchV)
2000/76/EC (17. BImSchV)
27. BImSchV
30. BImSchV
TA-Luft (Prevention of Air Pollution)
EN 15267
EN 14181
- Electrical safety: CE
- Mounting:
Mounting flange, DN125, PN6
Mounting flange, ANSI, 5″
- Test functions: Automatic check cycle for zero and span point (only for NH3 and HCl)
- Options: SCU control unit (for non-hazardous areas only)