MCS100FT system
- Measured values: CH4, CO, CO2, Corg, HCl, HF, H2O, NH3, NO, NO2, N2O, O2, SO2, NOx, C3H8, C2H6
- Performance-tested measurands: CH4, CO, CO2, Corg, HCl, HF, H2O, NH3, NO, NO2, N2O, O2, SO2
- Measurement principles: FTIR spectroscopy, flame ionisation detection, Zirconium dioxide sensor
- Gas flow rate: ≤ 300 l/h
- Measuring ranges:
CH4: 0 … 70 ppm / 0 … 210 ppm
CO: 0 … 60 ppm / 0 … 1,200 ppm
CO2: 0 … 25 Vol.-%
Corg: 0 … 7.5 ppm / 0 … 75 ppm
HCl: 0 … 10 ppm / 0 … 100 ppm
HF: 0 … 3 ppm / 0 … 10 ppm
H2O: 0 … 40 Vol.-%
NH3: 0 … 13 ppm / 0 … 65 ppm
NO: 0 … 150 ppm / 0 … 1,500 ppm
NO2: 0 … 25 ppm / 0 … 250 ppm
N2O: 0 … 25 ppm / 0 … 250 ppm
O2: 0 … 21 Vol.-%
SO2: 0 … 25 ppm / 0 … 525 ppm
NOx: 0 … 100 ppm / 0 … 1,000 ppm
C3H8: 0 … 25 ppm
C2H6: 0 … 40 ppm
Other measuring ranges and components on request
- Certified measuring ranges:
CH4: 0 … 50 mg/m³ / 0 … 150 mg/m³
CO: 0 … 75 mg/m³ / 0 … 300 mg/m³ / 0 … 1,500 mg/m³
CO2: 0 … 25 Vol.-%
Corg: 0 … 15 mg/m³ / 0 … 50 mg/m³ / 0 … 150 mg/m³ / 0 … 500 mg/m³
HCl: 0 … 15 mg/m³ / 0 … 90 mg/m³ / 0 … 150 mg/m³
HF: 0 … 3 mg/m³ / 0 … 10 mg/m³
H2O: 0 … 40 Vol.-%
NH3: 0 … 10 mg/m³ / 0 … 50 mg/m³
NO: 0 … 200 mg/m³ / 0 … 400 mg/m³ / 0 … 2,000 mg/m³
NO2: 0 … 100 mg/m³ / 0 … 500 mg/m³
N2O: 0 … 50 mg/m³ / 0 … 500 mg/m³
O2: 0 … 21 Vol.-%
SO2: 0 … 75 mg/m³ / 0 … 300 mg/m³ / 0 … 1,500 mg/m³
- Response time (t90): ≤ 200 s
FID: ≤ 45 s
- Sensitivity drift : < 3 % of the measuring range full scale value per maintenance interval
FID: < 2 % of measuring range full scale per week
- Zero point drift: < 3 % of the measuring range full scale value per maintenance interval
FID: < 2 % of measuring range full scale per week
- Detection limit : < 2 % of measuring range full scale
- Process temperature: ≤ +1,300 °C
- Sample gas temperature: ≤ +220 °C
- Process pressure: 900 hPa … 1,100 hPa
- Ambient temperature: +5 °C … +35 °C
- With cooling device: +5 °C … +50 °C
- Storage temperature: –20 °C … +60 °C
- Ambient pressure: 900 hPa … 1,100 hPa
- Ambient humidity: ≤ 80 %, Non-condensing
- Conformities:
Approved for plants requiring approval
2001/80/EC (13. BImSchV)
2000/76/EC (17. BImSchV)
27. BImSchV
TA-Luft (Prevention of Air Pollution)
EN 15267
EN 14181
MCERTS
- Electrical safety: CE
- Enclosure rating: IP43/ IP54
- Analog outputs: 0/4 … 22 mA, + 500 Ω, Number depends on system configuration; electrically isolated; max. 32 outputs
- Analog inputs: 0/4 … 22 mA, 100 Ω, Number depends on system configuration; electrically isolated; max. 32 inputs
- Digital outputs: + 48 V AC, 0.5 A, 35 W / 48 V DC, 0.5 A, 24 W, Number depends on system configuration; electrically isolated; max. 64 outputs
- Digital inputs: + 3.9 V, 4.5 mA, 0.55 W, Number depends on system configuration; max. 64 inputs
- Communication interface:
Modbus
PROFIBUS DP
Ethernet
- Type of fieldbus integration: TCP/ RTU RS-485
- Function: Connection to SOPAS ET software or OPC server
- Number: 1
- Indication:
LC display
Status LEDs: “Power,” “Failure,” and “Maintenance request”
- Input: Touchscreen
- Operation: Via LC-display or software SOPAS ET/ Several operating levels, password-protected
- Dimensions (W x H x D):
Analyzer cabinet: 806 mm x 2,165 mm x 605 mm (for details see dimensional drawings)
With cooling device: 1,096 mm x 2,165 mm x 605 mm
- Weight: ≤ 260 kg
- Power supply:
Power consumption: Analyzer cabinet ≤ < 1,000 W
- Without cooling device:
Sample gas line, heated: ≤ 95 W/m
Gas sampling unit: ≤ 450 W
Heated probe tube: ≤ 450 W
- Auxiliaries:
Zero gas (FTIR/FID): ≤ 300 l/h … ≤ 350 l/h, Instrument air; 3±0.2 bar; particle size max. 1 µm; oil content max. 0.1 mg/m³; pressure dew point max. –30 °C
Zero gas (O2) ≤ 350 l/h, O2 in N2: 1 … 4 Vol.-%; accuracy ±2%; 3 ±0.2 bar
Reference gas (FTIR): ≤ 350 l/h, Measuring component in N2; 70% of measuring range full scale; 3 ±0.2 bar
Reference gas (O2): ≤ 350 l/h, Ambient air
Reference gas (FID): ≤ 450 l/h, Propane in synthetic air: 75% of measuring range full scale value
Instrument air ≤ 2,000 l/h, Instrument air: 5 … 7 bar; particle size max. 1 µm; oil content max. 0.1 mg/m³; pressure dew point max. –30 °C
Fuel gas ≤ 4.8 l/h, Hydrogen: 5.0 or higher; 3 ±0.2 bar
Combustion air: ≤ 30 l/h, Instrument air; 3±0.2 bar; particle size max. 1 µm; oil content max. 0.1 mg/m³; pressure dew point max. –30 °C
- Sample gas connections:
Sample gas inlet: DN 4/6
Exhaust gas outlet: DN 8/10
- Auxiliary gas connections:
Test gas: DN 4/6
Fuel gas: DN 4/6
Instrument air: DN 6/8
- Corrective functions: Internal adjustment unit (option)
- Test functions: Internal zero point check
- Options: Integrated total hydrocarbon analyzer