DUSTHUNTER SB100 system
- Measured values: Scattered light intensity, dust concentration (after gravimetric comparison measurement)
- Performance-tested measurands: Scattered light intensity
- Measurement principles: Scattered light backward
- Spectral range : 640 nm … 660 nm, Laser, protection class 2, power < 1 mW
- Measuring ranges Dust concentration: 0 … 10 mg/m³ / 0 … 200 mg/m³, Higher measuring ranges on request
- Certified measuring ranges Scattered light intensity: 0 … 100 SI / 0 … 15 SI / 0 … 50 SI / 0 … 200 SI
- Response time (t90): 1 s … 600 s, Freely adjustable
- Accuracy: ≤ 2 %, Of measuring range full scale
- Process temperature: –40 °C … +600 °C
- Process pressure:
With MCU-P control unit: –50 hPa … 2 hPa
With external purge air unit: –50 hPa … 30 hPa
- Process gas humidity: Non-condensing
- Duct diameter: ≥ 500 mm
- Conformities: Approved for plants requiring approval
2001/80/EC (13. BImSchV)
2000/76/EC (17. BImSchV)
27. BImSchV
TA-Luft (Prevention of Air Pollution)
EN 15267
EN 14181
MCERTS
2010/75/EU
- Electrical safety: CE
- Test functions:
Automatic self-test (linearity, contamination, drift, aging)
Contamination limit values: warning at 20%, fault at 30%
Manual linearity test with reference filter
DHSB-T10/-T11 sender/receiver unit
- Ambient temperature: –40 °C … +60 °C
- Electrical safety: CE
- Enclosure rating: IP66
- Dimensions (W x H x D): 265 mm x 274 mm x 692 mm (for details see dimensional drawings)
- Weight : ≤ 10 kg
- Power supply:
- Voltage: 24 V
- Voltage: Supply via control unit
- Power consumption: ≤ 4 W
MCU-N control unit
- Description: Unit to control the system components and to evaluate and output the data provided by them
- Ambient temperature: –40 °C … +60 °C
- Electrical safety: CE
- Enclosure rating: IP66
- Analog outputs: 3 outputs, 0/2/4 … 20 mA, + 750 Ω, Electrically isolated; two additional outputs if using I/O modules (option)
- Analog inputs: 2 inputs, 0 … 20 mA, Not electrically isolated; two additional inputs if using I/O modules (option)
- Digital outputs : 5 relay contacts, + 48 V, 1 A, Potential-free; for status signals
- Digital inputs: 4 potential-free contacts
- Communication interface: Modbus/ PROFIBUS DP/ Ethernet
- Type of fieldbus integration:
TCP (via optional interface module, only one module possible per MCU)
RTU RS-485 (via optional interface module, only one module possible per MCU)
Via optional interface module (only one module possible per MCU)
- Function: Connection to SOPAS ET software or OPC server
- Number: 1
- Indication: LC display
- Status LEDs: “Power,” “Failure,” and “Maintenance request”
- Operation: Via LC-display or software SOPAS ET
- Dimensions (W x H x D): 210 mm x 340 mm x 120 mm (for details see dimensional drawings)
- Weight : ≤ 3.7 kg
- Power supply:
Voltage: 90 … 250 V
Voltage: Version with 24 V DC available as an option
Frequency: 47 … 63 Hz
Power consumption: ≤ 15 W
- Options:
Interface module(s)
I/O module(s)
MCU-P control unit
- Description: Unit to control the system components and to evaluate and output the data provided by them. With integrated purge air unit.
- Gas flow rate: ≤ 20 m³/h
- Ambient temperature: –40 °C … +45 °C, Intake temperatures for purge air
- Electrical safety: CE
- Enclosure rating: IP66
- Analog outputs: 3 outputs, 0/2/4 … 20 mA, + 750 Ω, Electrically isolated; two additional outputs if using I/O modules (option)
- Analog inputs: 2 inputs, 0 … 20 mA, Not electrically isolated; two additional inputs if using I/O modules (option)
- Digital outputs : 5 relay contacts, + 48 V, 1 A, Potential-free; for status signals
- Digital inputs: 4 potential-free contacts
- Communication interface: Modbus/ PROFIBUS DP/ Ethernet
- Type of fieldbus integration:
TCP (via optional interface module, only one module possible per MCU)
RTU RS-485 (via optional interface module, only one module possible per MCU)
Via optional interface module (only one module possible per MCU)
- Function: Connection to SOPAS ET software or OPC server
- Number: 1
- Indication: LC display
- Status LEDs: “Power,” “Failure,” and “Maintenance request”
- Operation: Via LC-display or software SOPAS ET
- Dimensions (W x H x D): 300 mm x 455 mm x 220 mm (for details see dimensional drawings)
- Weight : ≤ 13.5 kg
- Power supply:
Voltage: 90 … 250 V
Voltage: Version with 24 V DC available as an option
Frequency: 47 … 63 Hz
Power consumption: ≤ 70 W
- Options:
Interface module(s)
I/O module(s)
SLV4-2 purge air unit, 2BH1300, 3-ph
- Description: Unit to provide dust-free air for flushing of optical surfaces
- Gas flow rate: 38 m³/h … 63 m³/h, At 30 hPa counter pressure, depending on low pressure inside the filter
- Ambient temperature: –40 °C … +55 °C
- Electrical safety: CE
- Enclosure rating: IP54
- Dimensions (W x H x D): 550 mm x 550 mm x 257 mm (for details see dimensional drawings)
- Weight : 18 kg
- Power supply: Three-phase current
Δ: 200 … 240 V, 50 Hz, 2,6 A, 400 W
Y: 345 … 415 V, 50 Hz, 1,5 A, 400 W
Δ: 200 … 275 V, 60 Hz, 2,6 A, 500 W
Y: 380 … 480 V, 60 Hz, 1,5 A, 500 W
Δ: 270 … 330 V, 50 Hz, 2,0 A, 400 W
Y: 465 … 570 V, 50 Hz, 1,16 A, 400 W
Δ: 290 … 360 V, 60 Hz, 2,1 A, 500 W
Y: 500 … 600 V, 60 Hz, 1,26 A, 500 W
Δ: 230 V, 50 Hz, 2,7 A, 370 W
Δ: 115 V, 60 Hz, 3,0 A, 450 W
Δ: 220 … 270 V, 50 Hz, 2,5 A, 400 W
Y: 380 … 465 V, 50 Hz, 1,45 A, 400 W
Δ: 240 … 290 V, 60 Hz, 2,6 A, 500 W
Y: 415 … 500 V, 60 Hz, 1,55 A, 500 W
- Auxiliary gas connections: Purge air 40 mm
- Test functions: Pressure switch (switching point –35 hPa)
- Integrated components: 2-step air filter, type Europiclon, dust capacity 200 g