- Input: 3-wire Pt100 α=0.00385 IEC 60751
- Output: 4-20 mA
- Sensor break: According to NAMUR NE43
- Sensor short-circuit: According to NAMUR NE43
- Measuring range: -200…+850 °C / -328…+1562 °F
- Adjustments – Zero: Any value within range limits
- Adjustments – Minimum span: 20 °C / 36 °F
- Error compensation: Sensor and system error correction
- Ambient temperature: –40…+85°C / -40…+185°F
- Humidity: 0…98% RH (non-condensing)
- Vibration: Acc. to IEC 60068-2-6:2007, test Fc, 10 to 2000 Hz, 10 g
- EMC: Acc. to EN 61326-1 and EN 61326-2-3/ Within requirements for measured span except surge: standard deviation 1% of span.
- Accuracy: Max of ±0,1K or ±0,1% of span
- Long-term stability: Max. drift ±0,1% of span / year
- Adjustable filtering level: 0,13 to 54 s
- Galvanic isolation: No
- Ex-classifications (only for IPAQ C202X): ATEX: II 1G Ex ia IIC T6…T4 Ga/ IECEx: Ex ia IIC T6…T4 Ga
- Power supply:
Standard version: 6…32 VDC
Ex-version: 8…30 VDC
- Runtime counter: Hour counter for elapsed operational time
- Simulated output: Fixed current ouput during a maximum time of 15 min